Reference Ranges

Prolis employs 4 types of Reference Ranges, a test result is evaluated against when the result is either entered
manually or is transferred through the interface from an analyzer and applied. The evaluation outcome appears
as a Flag in front of the result. The evaluation is performed in the following order.
 
Client Custom Ranges
Prolis enables the laboratory maintain its each client with the client's specific ranges, basing on the age and gender.
If the system possesses the range setup for the client, for the test, the result of which is being evaluated, Prolis
will use it and stop the evaluation process. It is the priority# 1 range type.
The range format is as follows'
Test ID     Age From          Age To          Gender        Value From          Value To               Flag                    Behavior
1115               0                         6               Male                    0                         8.50                    Very Low             Panic
1115               0                         6               Female               0                         6.80                    Very Low              Panic
1115               7                        18               Male                  8.51                    19.19                     Low                     Caution
1115               7                        18              Female              6.81                    18.00                     Low                     Caution
748                 0                       130              Male               Non-Reactive        Non-Reactive        Normal                Ignore
748                 0                       130              Male                 Reactive               Reactive               Abnormal           Panic
748                 0                        130            Female           Non-Reactive          Non-Reactive      Normal                 Ignore
748                 0                       130              Female             Reactive              Reactive               Abnormal             Caution
-------------------
This range type accommodates both qualitative and quantitative tests.
Use these ranges thoughtfully, covering the entire range from lowest to highest and the age from 0 to 130 years.
 
AG System Ranges
This age and gender based range can be maintained for a quantitative test.
If the system possesses the AG range for the test, the result of which is being evaluated, Prolis will use it and stop the evaluation process. It is the priority# 2 range type.
The range format is as follows'
Age From     Age To          Gender          Value From          Value To               Flag               Behavior
0                         6               Male                    0                         8.50                    Very Low             Panic
0                         6               Female               0                         6.80                    Very Low              Panic
7                        18               Male                  8.51                    19.19                     Low                     Caution
7                        18              Female              6.81                    18.00                     Low                     Caution
-------------------
-------------------
Use these ranges thoughtfully, covering the entire range from lowest to highest and the age from 0 to 130 years.
 
N System Ranges
This numerical range can be maintained for a quantitative test.
If the system possesses the this range for the test, the result of which is being evaluated, Prolis will use it and stop the evaluation process. It is the priority# 3 range type.
The range format is as follows'
Value From     Value To          Flag                    Behavior
0                         8.50                 Very Low             Panic
8.51                    19.19               Low                     Caution
19.20                  60.60               Normal                Ignore
60.61                 105.10              High                    Caution
105.11               50000.00          Very High            Panic
                
C System Ranges
This Choice based range can be maintained for a qualitative test.
If the system possesses this range for the test, the result of which is being evaluated, Prolis will use it and stop the evaluation process. It is the priority# 2 range type.
The range format is as follows'
Choice          Flag          Behavior
Negative        Normal         Ignore
Positive         Abnormal     Caution
Detected       Critical          Panic
 
Note:   Panic behavior issues alerts within Prolis, executing some routines for compliance.